1 Notes on commissioning


Please read the chapters of this manual carefully in the order given in order to learn how to use the device correctly.

 

Before you can carry out a measurement, you must download the VITALITY CHECK app to your smartphone or tablet (chapter 5) and charge the battery of your device (chapter 8). When the battery is low, it is impossible to turn on the device successfully.

 

Chapter 9 describes all the steps you need to take for a measurement. When the device is switched on, the light of the indicators always informs you of the current status. If the display does not match the descriptions in Chapter 9, you have not followed the current step correctly. In this case, please read Chapter 11 ("Troubleshooting") for the information on the corresponding display. For each display, a method is described how you can exit this state and continue with the measurement. Since the lighting of the displays has an automatic timeout, please note the configuration of the lighting before you search in Chapter 11.

 

 

2 Security

Read the safety instructions carefully before using the VITALITYCHECK.

Keep these instructions for future reference.

 

Use the device only as described in the instructions

Do not look into the light with which the device is measuring. Keep your hand on the sensor during the entire measurement. Staring into the active lights can cause permanent or non-permanent damage to your eyes.

 

Do not use the device near water or spill liquid on the device, as this could cause a short circuit, fire, or electric shock, which could result in serious injury, death, and property damage.

 

Do not use or store the device in extreme temperatures (e.g. do not leave it in a car with direct sunlight). The device could overheat, pose a risk of burns and fire and stop working.

 

Do not drop or drop objects on the device, as this could damage the battery, which could result in an explosion that could result in serious injury, death, and property damage.

 

Do not throw the device into a fire as this can cause an explosion, which can result in serious injury, death, and property damage.

 

Do not use your device on an airplane because the Federal Aviation Administration (FAA) and Federal Communications Commission (FCC) regulations prohibit the use of wireless devices in the air. Turn off your device before boarding an aircraft. Using the device on an airplane may affect airplane instruments, communications, and performance, disrupt the network, jeopardize the operation of the airplane, crew, and passengers, and may be illegal.

 

Do not dismantle the device.

 

Most modern electronic devices are shielded from radio frequency (RF) signals. However, certain electronic devices may not be shielded from the RF signals from the device. Keep the device away from medical equipment such as B. pacemakers and hearing aids, as they malfunction and can seriously injure or kill the user or other people.

 

Do not use the device in healthcare facilities as there may be devices that are sensitive to external RF energy.

 

The VITALITY CHECK contains an internal, non-replaceable lithium-ion battery.

Use only certified USB Type-C ™ cables and suitable USB chargers to charge or power the product.

 

Do not charge the device in a high humidity environment.

Do not cover the device or charger while it is charging.

 

 

Do not dispose of the device with unsorted waste. Improper disposal can be harmful to the environment and human health. Please contact the local waste authority for information about return and collection systems in your region. This device must not be disposed of with normal household waste. Instead, it should be disposed of by returning it to the point of sale or a municipal recycling collection point.

CAUTION!!

RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN INCORRECT TYPE. DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS


3 Intended use

The VITALITY CHECK may only be used for the purpose stated in these instructions for use.

 

 

4 system requirements

USB charger (can be a wall charger, a car charger or the USB port on a PC. The port must deliver a current of at least 500 mA)

2.4 GHz WiFi connection, at least 54 MBps

 

Internet connectivity

iPhone with IOS 12 or newer or Android phone with OS version 8 or newer

 

 

5 Download the VITALITY CHECK app

In order to use the VITALITY CHECK, the app for IOS and Android must be downloaded to your smartphone or tablet. It will soon be available in the Google Play Store and Apple App Store

 

 

6 Package contents

VITALITY CHECK device

Quick start guide

 

 

7 Product description

The VITALITY CHECK device enables the non-invasive measurement of biomarkers in human skin and blood. The technological basis is the MSRRS measurement (Multiple Spatially Resolved Reflection Spectroscopy). MSRRS enables optical detection of inhomogeneously distributed biomarkers in different depths of the tissue when the user places his hand on the sensor.

 

The device consists of a lid and the base. To open the device, hold the base and turn the lid counterclockwise.

 

The basis of the VITALITY CHECK is shown in (Figure 2). The base consists of an on / off button (2), a connectivity button (4), the USB-C socket (3), the sensor area (9) and several displays (5, 6, 7).

Requirements
Download
Description
Bild1.png

 

  • On / Off button (2): switch on when pressed, hold down to force switch off.

  • USB-C socket (3) can be used to charge the device.

  • Connectivity button (4): hold down to enter the initialization mode. For further details see chapter 9.2)

  • The status indicator (5) and can light up in almost any color and is used to indicate certain states. For more information, see below.

  • To show the progress of the measurement, the progress indicator (6) consists of three characters and lights up green during the measurement. The three symbols turn on one after the other until all symbols light up. When the measurement is finished, the light goes out. See below for more details.

  • The WiFi indicator (7) lights up green and shows the connection status of the device. See below for more details.

  • Groove (8) for aligning the hand when measuring.

  • When measuring, the LEDs of the sensor area (9) are switched one after the other

The LED of the status display (5) lights up to indicate the current status. The LED colors provide different basic information:

 

  • Red color: bug or problem

  • Orange color: not fully operational

  • Green color: ready for use

  • Blue color: initialization mode. In this mode the scanner opens a WiFi network (see chapter 9.2). Connect to this network to make communication settings.

  • Color cyan: update mode

  • White: USB connected, disconnect USB to enter measurement mode

 

 

Progress indicator (6) shows the progress. The information displayed has a different meaning for each status. See Chapter 11 for more information.

 

 

The WiFi indicator (7) can have the following status:

 

 

Not lit: The scanner is not connected to a WiFi network that is an existing router hotspot

 

  • Green color: The scanner is connected to a WiFi network

  • Flickering: The scanner is sending data to a WiFi network

  • Blinking: The scanner is about to connect to a WiFi network

8 Charging the Battery

 

Any charger that is equipped with a USB Type-C ™ connector is suitable for charging the device. As soon as a charger is connected, the battery is charged and the large icon in the progress indicator lights up red. When the battery is charged, the device will automatically stop charging and the red light will turn off.

Battery

Please charge the VITALITY CHECK device before using it for the first time

​

  • Plug the USB-CTM cable into the charging port

​

  • Connect the cable to a USB charger or a USB port

​

  • A red LED lights up during the charging process. The LED is off when the battery is charged. The LED is located in the largest recess of the progress indicator / Figure 3

​

  • The device should be disconnected after approx. 3 hours

​

  • The progress indicator lights red to indicate that the device is charging

Bild7.png
Bild8.png

When the VITALITY CHECK is switched off, no LED lights up. If an LED lights up, please first switch off the device by pressing the ON / OFF button for a long time (approx. 15 seconds).

​

When you turn on the device by pressing the ON / OFF button, the status indicator should light up blue. If not, please press the connection button until the status indicator lights up blue. Blue indicates that the scanner has opened a WiFi access point (Figure 4). How to connect the device to the WiFi access point is described in chapter 9.2.

​

If there is no further action after a timeout of 60 seconds, the device switches off automatically.

9 steps to take a measurement

 

In order to successfully carry out a measurement with the VITALITY CHECK, please follow the steps below in the order described.

Measurement

It is not possible to carry out a measurement and to charge the battery at the same time, as charging can affect the quality of the measurement. Please remove the USB cable before starting a measurement.

9.1 Switching the device on and off

 

If you turn on the scanner by pressing the power button while the battery is charging, or if the battery has previously been charged and the scanner is still connected to the USB cable, the status indicator will glow white (Figure 5) and will not open a WiFi access point. Please disconnect the USB cable.

On OFF
Unbenannt.PNG

If the device's battery level is too low to take a measurement, the status indicator will flash red several times before the device automatically turns off. If the device does not turn on, the battery may be completely empty. In both cases, please charge the battery (Chapter 8).

Bild11.png
connect

9.2 How to establish a connection with the access point / WiFi of the scanner

 

 

After you have successfully switched on the device, the status indicator lights up blue. If not, press the connect button until the indicator glows blue. Blue indicates that the scanner has opened a WiFi access point. In access point mode, the scanner provides WiFi that a smartphone can connect to.

 

 

Now switch on your smartphone / tablet, open the Vitality app and select the "Setup wizard" by pressing the three-bar symbol in the top left (Figure 7). Please follow the procedure of the "setup wizard" step by step to connect the scanner and phone / tablet (chapter 9.3 and 9.4) and then to connect the scanner to the external WiFi or to the hotspot of a phone / tablet (chapter 9.5).

 

 

In order to enable a connection between the scanner and the phone / tablet, the wizard asks you to open the settings of your phone / tablet and the selection of the WiFi settings. All WiFi networks in the area are displayed. Please select biozoom_xxxx (Figure 7), this is the WiFi name of the access point of your scanner. A four-digit number consisting of numbers (0-9) and / or letters (af) is displayed in place of the "xxxx". The specific WiFi name of your scanner is printed on the back of the scanner. Some phones / tablets inform you that the selected WiFi has no internet connection. This is correct because the access point only connects the scanner and the phone / tablet. In some cases the phone / tablet will ask you to confirm that it is not connected to the internet. Please confirm this. If you don't confirm, the scanner and phone / tablet will be connected.

Bild1.png

9.3 Confirmation of a connection with the access point

 

 

The smallest symbol in the progress indicator (6) lights up green when the scanner is connected to a telephone.

Bild1.png

9.4 Confirming a connection between app and scanner

The first and second icons of the progress indicator light up green and the status indicator lights up blue when the scanner is connected to the VITALITY CHECK app on your smartphone/tablet.

Bild2.png

9.5 Selecting the WiFi for historical measurements

When you perform a measurement, the measured data is sent to a central server for calculation. The result is displayed on the phone / tablet. Therefore, the scanner must be connected to a WiFi network that transmits the data to the central server via the Internet. There are two ways to do this:

​

Use of an existing WiFi, e.g. B. Your home WiFi or office WiFi

Apply the phone / tablet hotspot

​

If you have not yet set up a hotspot on your phone, you must first configure it accordingly, as the name and password of the hotspot must be known to connect the scanner and hotspot. Open your phone's settings. In most cases, the hotspot is a submenu of the WiFi settings. Use a name and password that you can easily remember for configuring the hotspot; it will help you connect the scanner to the hotspot.

The following existing WiFi's cannot be used:

  • WiFi networks that require you to enter information into a browser to gain access, such as: B. Hotel WiFi's, WiFi on some trains or planes

  • WiFi networks that do not allow communication from one device to another, e.g. B. Hotel WiFi's, company guest networks

In all these cases, please select the hotspot on your phone / tablet as an alternative

wifi

The scanner continuously scans for WiFi networks and sends the networks it finds to the phone / tablet via the access point (see Section 9.4). The networks are displayed in the VITALITY CHECK app. Please select the desired WiFi network and enter the password in the opened window. If the scanner can connect to the specified WiFi network, all three symbols of the progress indicator light up green. The status indicator lights up blue.

Bild3.png

When the scanner is connected to the WiFi network, the access point will be closed. Therefore the WiFi indicator and the status indicator light up green and the light of the progress indicator is switched off.

Bild4.png

In order to continue to enable communication between the scanner and the phone / tablet, the phone / tablet must now also be connected to the same WLAN that the scanner is connected to. If this WLAN is your home WLAN or your office WLAN, your phone / tablet was usually connected to this WLAN beforehand. In this case, your phone / tablet automatically connects to this WLAN and no further action is necessary to connect the scanner and phone / tablet.

In all other cases you have to open the settings of your phone / tablet. Please select the submenu for the WiFi settings. Connect your phone to the same WiFi network that the scanner is connected to.

The scanner and phone / tablet are now communicating via the router (existing WiFi network).

The VITALITY CHECK is now ready for the measurement. The app displays the page for starting a measurement.

Bild5.png

9.5.2 Set up the hotspot of the phone / tablet

 

You have successfully followed the instructions in chapters 9.2 - 9.5. Therefore, the status indicator lights up blue and two icons of the progress indicator light up green, indicating that the scanner and the app are connected via the access point of the scanner. The scanner now continuously scans for WiFi networks, the networks found are displayed in the app.

 

 

In order to connect the scanner to the hotspot of the phone / tablet, you have to press the plus sign (+) at the top of the screen. In the window that now opens you can enter the SSID (name of the hotspot) and the password of the hotspot. Press the word "Connect" to complete the process.

hotspot
Bild6.png

The information is sent to the scanner via the existing access point connection. The attempt to connect the scanner to the hotspot must of course fail because the hotspot has not yet been opened.

Bild7.png

Therefore please open the WiFi settings of the smartphone and switch on the hotspot (Figure 13a). The connection to the access point is then terminated. Because the scanner saved the credentials it sent, all of the progress bar characters will turn on to confirm that the hotspot is now open.

Bild8.png

Please switch the scanner off and on again. The scanner is automatically connected to the hotspot. To indicate this, the status indicator and the WiFi indicator light up green and the light of the progress indicator is off.

The VITALITY CHECK app displays the page for starting a measurement (Figure 14). You can start a measurement by pressing the apple symbol.

Bild9.png

9.5.3 Automatically connect to a WiFi network

 

 

The scanner saves the login details of the last connected WiFi network. Therefore, the next session will automatically connect to this network if this network is available. This is very convenient as it allows you to skip some of the steps described above. Please note that only the last network used is saved. However, there are small differences in how you connect to an existing one

 

WiFi network or a hotspot.

 

Connection to an existing WiFi network

 

For an automatic connection, please proceed as follows:

 

 

Turn on your phone / tablet

 

 

Since your phone / tablet saves the access data of the WiFi networks to which you were previously connected, the connection is established immediately after switching on. If your phone / tablet was already switched on, the connection will of course be established if the router is visible to your phone / tablet.

 

Note: If you have multiple routers or WiFi access points, you will need to connect your phone to the same WiFi that your scanner is configured for.

 

 

Open the VITALITY CHECK app

Turn on the scanner

 

 

The status indicator lights up orange briefly, but immediately changes to green. The WiFi indicator also lights up green. As the scanner is now connected to the existing WLAN, it is ready to carry out a measurement.

Bild10.png

Connection to the hotspot

For an automatic connection, please proceed as follows:

  • Turn on your phone / tablet.

  • Now open the WiFi settings of your phone / tablet. There you will find a submenu for opening a WLAN hotspot. Please activate the hotspot.

  • Turn on the scanner


The status indicator briefly lights up orange, but immediately changes to green. The WiFi indicator also lights up green. Now that the scanner is connected to the hotspot, it is ready to take a measurement.

Bild11.png

9.6 Performing a measurement

Before measuring for the first time, please read the general recommendations for correct measurement, which are listed in Chapter 10 "Measurement Basics". It will help you a lot to get a correct result.

9.6.1 Automatic hand recognition

 

You have opened the VITALITY CHECK app and see the "Start measurement" button (status display and WiFi display light up). Please press the button now. The smallest symbol on the progress indicator flashes green and prompts you to place your hand on the sensor.

 

Note: Please carefully read the instructions on how to place your hand on the sensor in order to obtain a correct measurement (see chapter 10).

 

 

The scanner will now check to see if your hand is present. Every time the scanner looks for your hand, it will blink with orange LEDs. The measurement is only started when the hand is on the sensor. This is a safety mechanism to prevent the user from being disturbed by the bright light of a measurement in progress. An automatic timeout stops the orange flashing after 60 seconds if the user has not placed his hand on the sensor.

Palm
Bild12.png

9.6.2 Automatic start of the measurement

When the sensor has recognized that the hand has been placed on, the measurement begins and the smallest symbol in the progress indicator stops flashing and lights up green continuously. At this point, the hand should not be moved on the sensor to avoid incorrect measurements. The small symbol indicates that the measurement is in the first half.

​

The measurement data is continuously transmitted to the server for calculation via WiFi and the Internet. Every time a data packet is sent, the WiFi indicator will flicker.

Measurement
Bild13.png

9.6.3 Measurement progress

 

Since the measurement is running, the first and second symbols of the progress display light up after a while.

Bild14.png

The measurement is finished when all three symbols of the progress indicator light up green. The hand should now be removed from the sensor.

Bild15.png

If you do not remove your hand, the smallest symbol in the progress indicator will start to flash. Please remove your hand now.

Bild16.png

When you have removed your hand, the first two symbols of the progress indicator are switched off immediately. The third symbol lights up until all data has been transferred to the server.

Bild17.png

When all data has been sent to the server, the largest symbol of the progress indicator is also switched off and only the status indicator and the WiFi indicator remain lit.

Bild18.png

If you are taking an antioxidant-intensive measurement, the next measurement must be taken. Therefore the smallest symbol of the progress indicator flashes again and four LEDs of the sensor switch on continuously one after the other. It indicates that you need to place your hand on the sensor surface again. Please follow the procedure described in the previous chapters (9.6.1 ff). If you do not hang up your hand, an error message will appear on the phone / tablet after 60 seconds.

Bild19.png

When you have carried out all four measurements of the intensive measurement method or have only carried out a single measurement, the scanner is automatically switched off. At the same time, the app informs you that the result is available. If you click on the link that is displayed, a website will open that presents the result and recommendations for improving it.

If you want to take another measurement, you have to open the app and turn on the scanner again.

Bild20.png
Measurement basics

10 Basics of measurement

 

The VITALITY CHECK is a sensitive sensor that is able to detect certain substances on the human skin. It is helpful to follow the tips below for accurate and reliable results.

 

 

Preparation before starting the measurement

 

Rest your body before starting the measurement.

 

 

Exercising immediately before the measurement often has an impact on the antioxidant measurement as well as heart rate variability that doesn't reflect your general level. If you want maximum measurement accuracy, you should give your body 20 minutes of rest between training and the measurement.

 

 

Measure at the same time each day for consistent results. Antioxidant levels can change throughout the day.

 

 

Your body relaxes and recovers while you sleep. You can see that in the antioxidant level too. Most people have a higher value in the morning and during the activity and stress of the day the value goes down and is lowest in the evening. For some people, it increases right after a healthy meal and lunch break. The changes between morning and evening are typically on the order of 0.1 to 0.5. One can also find differences between Monday and Friday due to workload or occupational stress and weekend relaxation.

 

 

Measure only on dry skin. After you've washed your hands with soap and dried them thoroughly, water will remain on your skin for a while. It will have a little impact on the measurement, so it is better to wait 30 minutes after washing your hands for the most accurate results.

 

Measure only on uninjured, healthy, untreated, bare, and dry skin for consistent results. If you suffer from any skin condition it is likely that the sensor will not measure properly as it is only calibrated for healthy skin.

 

Skin treatments of any kind, e.g. B. skin care products, cosmetics, sun protection, tattoos influence the measurement. Hair also has an impact. We therefore recommend measuring at the root of the thumb of the right hand.

 

 

Performing the measurement

 

 

  • Measure while seated with the scanner on the table.

  • Place your elbow on the table to check the pressure when you place your hand on the sensor. Do not use extra force, the weight of your hand / arm will suffice

 

 

You can achieve the ideal contact pressure by letting your hand rest with its own weight on the sensor. Do not use the muscles of the arm to raise your hand or press the sensor. The ideal contact pressure corresponds to a weight of about 800 g, i.e. about two pounds. The reason for this is that your blood contains antioxidants. If you press too hard on the sensor, you will push some blood out of the skin and change the effective antioxidant level.

 

The sensor has a function that compensates for the effect of the contact force. So you don't have to worry about getting perfect contact pressure. But if your contact pressure is ideal, you will still get a slightly better result. Forcing blood out of the skin makes it completely impossible to measure heart rate variability.

Bild21.jpg
Bild22.jpg

Sitting position - sensor on the table. Place your elbows on the table - no additional force when measuring

​

  • Gently place your right hand on the sensor using the groove to precisely align the hand. The measurement always at the same point helps you to correctly recognize the effects of a change in diet and lifestyle.

​

  • Do not move your hand or shift your weight during the measurement.

​

It is important to keep your hand steady during the measurement. The sensor takes a number of different measurements and compares the results. If the hand moves between measurements or the contact with the sensor varies, the comparison works less well.

Bild23.jpg
Bild25.png
  • Avoid bright light on the hand during the measurement (sunlight, bright artificial light). Cover your hand if necessary.

Bright light can penetrate your hand. You can try this out by turning on your smartphone's LED flashlight and placing your finger on it. You can see that especially red light can penetrate the finger completely. Sunlight is even stronger and can penetrate your entire hand. This is why the biozoom sensor refuses to measure in direct sunlight. If you want maximum precision, you can help the sensor by making sure that no bright light hits the back of your hand during the measurement. Normal ceiling light is usually not a problem.

​

  • General aspects that affect the level of the measurement results

  • Note that prolonged exposure to the sun or using a tanning bed can temporarily lower your antioxidant levels.

​

Influence of wearing gloves

If you wear gloves to work you should be aware that sometimes there is a significant impact on antioxidant levels. You will get better results if you always measure before work.

​

  • Note that illness can lower your antioxidant levels.

​

​

When you have a cold or any other illness, it puts stress on your body. Be aware that when this happens, your antioxidant levels will usually go down. Wait until you have recovered before comparing two results.

​

  • Appropriate exercise will help your antioxidant levels, but vigorous exercise can temporarily lower your antioxidant levels.

  • Note that smoking results in consistently lower levels of antioxidants compared to not smoking

  • Note that excessive alcohol consumption can temporarily lower antioxidant levels

11 Troubleshooting


Below is a table that summarizes all the functions and displays relevant to using the scanner. If you encounter unexpected behavior on the part of the scanner, this table will help you understand it quickly. This table will inform you if further action is required.

 

In the event of unexpected events not listed in the table, please take the following steps:

 

 

  • Turn off the scanner by pressing and holding the ON / OFF button for 15 seconds. Restart the app.

  • Turn the scanner back on by pressing and holding the ON / OFF button for 1 second.

  • Because the scanner indicates its status by lighting up a specific color or a combination of colors of different indicators, the following table is ranked by color.

 

1 scanner

2 On / Off button

3 USB-C socket

4 connectivity button

5 Status display

6 progress indicator

7 WiFi indicator

8 groove

9 sensor range

Bild26.png
Troubleshooting
Maintenance
technical
certifications

13 Technical Specification

12 Maintenance and cleaning

 

The VITALITY CHECK does not require regular maintenance and has no user-serviceable parts.

 

Regular cleaning is recommended to ensure consistent measurement quality. Please use a slightly damp, soft cleaning cloth.

 

 

Never use paint thinner, benzene, alcohol, disinfectants or disinfectants or other strong cleaning agents to clean the VITALITY CHECK, as these can damage the housing and the optical surface of the sensor.

14 certifications

Bild27.png